OPERATING MANUAL 2945A
46882-311D 5-25
(1) Refer to Results table 5-41 on page 5-43. Connect the equipment as shown in Fig. 5-17,
connecting the incoming signal to the antenna input socket of the UUT. Lock the Service
Monitor external standard input to the external standard output of the signal generator.
(2) Set the UUT to [Tx TEST], RF IN/OUT [SELECT] antenna BNC input socket. Press
[HELP/SETUP], [Setup], and toggle RF counter resolution until 1 Hz is displayed in reverse
field, then press [Return], [Return] to get back to the TX test display.
(3) On the UUT, press [Tx Freq] 10 MHz. Set the RF signal generator to provide a signal of
10 MHz at an RF level of
60 dBm. Check this level by removing the RF test signal
connection from the UUT and applying it to the RF power meter. Adjust the level if
necessary, then reconnect the UUT.
(4) Check that the offset indicated on the UUT is 0 Hz
±
1 Hz.
(5) Repeat steps 3 & 4 for frequencies of 500 MHz and 1.05 GHz.
(6) Refer to Results table 5-42 on page 5-44. Press the [Auto Tune] key on the UUT.
(7) Set the RF signal generator to provide a signal of 1000 MHz at a level of
13 dBm (check
and set this level on the power meter). Check that the UUT autotunes to the incoming
frequency and the Frequency counter indicates the frequency
±
1 Hz. Repeat at 500 MHz and
10 MHz.
(8) Disconnect the external reference from the UUT and connect an external reference with an
accuracy of 1 part in 10
9
or better to the RF signal generator.
(9) Refer to Results table 5-43 on page 5-44. Set the RF signal generator to provide a frequency
of 1000 MHz at a level of 0 dBm.
(10) Check that the UUT indicates a frequency between 999.999500 MHz and 1000.000500 MHz.
If the instrument is fitted with option 3, the high stability (OCXO) internal frequency
standard, check that the reading falls between the limits 999.999750 and 1000.000250. The
test limits in this step are for guidance and assume that the internal frequency standard has
been recently adjusted. Ageing and stability have to be considered when establishing the real
test limits.
RF spectrum analyzer
Specification
Frequency range: 100 kHz to 1.0 GHz.
Resolution bandwidth 300 Hz, 3, 30, 300 kHz, 3 MHz.
On screen dynamic range: 80 dB.
On screen linearity: Typically ± 2 dB ± resolution (10 dB/div) (10 dB above the noise floor)
Resolution: 0.1 dB on 2 dB/ division
0.5 dB on 10 dB/ division
Level flatness ±1 dB ± resolution over 50 MHz span.
Test equipment
Description Minimum specification Example
RF signal
generator
100 kHz to 1 GHz. Level accuracy ±0.85 dB IFR 2041
RF power meter ±0.1 dB from 10 MHz to 1 GHz IFR 6960/A/B +
6912 sensor