OPERATING MANUAL 2945A
46882-311D 5-11
UUT
C2587
RF INPUT
BNC
OUTPUT
Modulation meter
Fig. 5-8 Internal AM accuracy checks
(1) Refer to Results table 5-10 on page 5-34. Connect the test equipment as shown in Fig. 5-8
(BNC output socket on UUT).
(2) Set the UUT to [Rx TEST], RF IN/OUT [SELECT] BNC output, [RF Gen],
[LEVEL] 0 dBm, [FREQ] 1.5 MHz, [Mod Gen], [Gen 2], [FREQ] 1 kHz, [LEVEL] 50%.
(3) Set the Modulation meter to monitor AM in a 0.3 to 3.4 kHz bandwidth.
(4) Check the Modulation meter for a reading within
±
7% of setting
±
1 digit.
(5) Repeat step 4, but with the UUT RF generator set to each of the remaining carrier frequencies
shown in the Results table 5-10 on page 5-34.
(6) Refer to Results table 5-11 on page 5-34. Set the UUT to [RF Gen], [FREQ] 100 MHz,
[LEVEL] 5 dBm, [
INC] 1 dB, [Mod Gen], [Gen 2], [LEVEL] 70% and check the
Modulation meter for a reading within
±
7% of setting
±
1 digit. Using the level decrement
arrow key, decrement the RF level by 1 dB and repeat check. Repeat in 1 dB decrements
down to
15 dBm.
(7) Refer to Results table 5-12 on page 5-35. Set the UUT to [RF Gen], [LEVEL] 0 dBm,
[Mod Gen], [Gen 2], [LEVEL] 5%.
(8) Check that the Modulation meter reads within
±
7% of setting
±
1 digit. Repeat for UUT AM
depths of 10, 20, 30, 40, 50, 60, 70, and 85%.
(9) Refer to Results table 5-13 on page 5-35. On the UUT, press [FREQ] 65 Hz to set Mod Gen
2 to 65 Hz.
(10) Set the Modulation meter to monitor AM in a 30 Hz to 50 kHz flat bandwidth.
(11) Check the Modulation meter for a reading within
±
10% of setting
±
1 digit.
(12) Repeat step 11 with the UUT set to each of the following mod rates:- 140 Hz, 500 Hz, 2 kHz,
and 5 kHz.
(13) Set the UUT mod rate to 15 kHz and check the Modulation meter for a reading within
±
15%
of setting
±
1 digit.
(14) Refer to Results table 5-14 on page 5-35. Reset Mod Gen 2 frequency to 1 kHz and level to
30%.
(15) Set the Modulation meter to monitor AM in a 0.3 to 3.4 kHz bandwidth.
(16) Connect the LF output from the Modulation meter to the AF input on the UUT. Select the
0.3 to 3.4 kHz AF Filter on the UUT and select distortion measurement.
(17) Check that the distortion indicated on the UUT display is less than 2 %.