OPERATING MANUAL 2945A
46882-311D 5-13
(10) Switch distortion measurement off.
(11) Refer to Results table 5-18 on page 5-36. Set the UUT Mod Gen 2 to provide 10 kHz
deviation level at a frequency of 50 Hz.
(12) Select the 30 Hz to 50 kHz filter on the Modulation meter.
(13) Check that the Modulation meter reads 10 kHz deviation
±
10%
±
10 Hz. Repeat for UUT
modulation frequencies of 10 kHz and 15 kHz.
AF output tests
Audio generator output level
Specification
Level range: 0.1 mV to 4 V RMS
Accuracy: ±5% ±1 resolution 50 Hz to 15 kHz
Test equipment
Description Minimum specification Example
DVM 1 % accuracy 50 Hz to 15 kHz Solatron 7150+
C2588
DVM
VOLTMETER
INPUT
TERMINALS
AF GEN
OUTPUT
UUT
Fig. 5-10 Audio generator level accuracy checks
(1) Refer to Results table 5-19 on page 5-37. Connect the test equipment as shown in Fig. 5-10,
connecting the UUT AF Gen output socket to the DVM voltmeter input terminals.
(2) Set the DVM to measure Volts AC RMS.
(3) Set the UUT to [AF TEST], [Audio Gen] 1 [FREQ] 1 kHz, [LEVEL] 4 V (ensure that AF
Gen 2 is OFF). Check that the DVM reads the level set
±
5%
±
1 mV. Repeat with
frequencies of 50 Hz, 150 Hz, 500 Hz, 2 kHz, 5 kHz, 10 kHz and 15 kHz in turn, checking
that the DVM reads the level set
±
5%
±
1 mV.
(4) Set the UUT to [AF Gen 1], [FREQ] 1 kHz, [LEVEL] 20.0 mV. Repeat with the level set to
400 mV. Check that for each level the DVM reads the level set ±5% ±0.1 mV. Repeat with
the level set, in turn, to 1000 mV, 2000 mV and 3000 mV, and check that in each case the
DVM reads the level set ±5% ±1 mV.
(5) Refer to Results table 5-20 on page 5-37. Repeat the above checks for AF Gen 2 with AF
Gen 1 switched OFF.